dc.contributor.author | Güngör, Murat Alparslan | |
dc.contributor.author | Karagöz, İrfan | |
dc.date.accessioned | 2019-05-13T09:08:58Z | |
dc.date.available | 2019-05-13T09:08:58Z | |
dc.date.issued | 2016 | |
dc.identifier.citation | Güngör, M. A., Karagöz, İ. (2016, June). The analysis of tissue temperature rise caused by linear array transducers with Program Spheresum. 2016 International Conference on Optoelectronics and Image Processing (ICOIP), Warsaw, Poland. | en_US |
dc.identifier.isbn | 9781509008797 | |
dc.identifier.uri | https://doi.org/10.1109/OPTIP.2016.7528519 | |
dc.identifier.uri | https://hdl.handle.net/11491/2081 | |
dc.description | 2016 International Conference on Optoelectronics and Image Processing, ICOIP 2016, 10 June 2016 through 12 June 2016, | en_US |
dc.description.abstract | Diagnostic ultrasounds transmit sound waves into the body, and thus reveal important information about it. Because the transmitted sound waves cause tissue heating, diagnostic ultrasounds can produce adverse biologic effects in tissue, namely, thermal bioeffect. Both tissue characters and ultrasound parameters affect the amount of temperature rise on the tissue. Program Spheresum [1] gives the temperature rise on the tissue scanned by a diagnostic ultrasound. This study compares tissue temperature rise due to linear array transducers with different parameters by using Spheresum. The results show that the maximum temperature rise is generally near the source surface in the linear array transducer application with high frequency for different focal points. Especially in tissue with a high absorption coefficient and for the linear array transducer with high frequency, tissue temperature rise near the source surface should be considered as the image resolution. © 2016 IEEE. | en_US |
dc.language.iso | eng | |
dc.publisher | Institute of Electrical and Electronics Engineers Inc. | en_US |
dc.relation.isversionof | 10.1109/OPTIP.2016.7528519 | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Diagnostic Ultrasound | en_US |
dc.subject | Temperature Rise | en_US |
dc.subject | Ultrasound Parameters | en_US |
dc.title | The analysis of tissue temperature rise caused by linear array transducers with Program Spheresum | en_US |
dc.type | conferenceObject | en_US |
dc.relation.journal | Proceedings - 2016 International Conference on Optoelectronics and Image Processing, ICOIP 2016 | en_US |
dc.department | Hitit Üniversitesi, Mühendislik Fakültesi, Elektrik Elektronik Mühendisliği Bölümü | en_US |
dc.authorid | 0000-0001-7446-7808 | en_US |
dc.identifier.startpage | 54 | en_US |
dc.identifier.endpage | 58 | en_US |
dc.relation.publicationcategory | Konferans Öğesi - Uluslararası - Kurum Öğretim Elemanı | en_US |