dc.description.abstractAmaç: Keratokonusta, Pascal dinamik kontur tonometre (DKT), Keeler Pulsair EasyEye non-kontakt tonometre (NKT) ve Goldmann aplanasyon tonometresi (GAT) ile yapılan göz içi basıncı ölçümleri karşılaştırılmıştır.Gereç ve yöntem: Çalışmaya 36 keratokonuslu olgunun 62 gözü alındı. Göz içi basıncı ölçümleri hastalar 5 dakika oturtulup dinlendirildikten sonra yapıldı. Her cihaz için en az 3 göz içi basıncı ölçümü alınıp ortalama değer kaydedildi. Tüm ölçümler aynı kişi tarafından aynı günde gerçekleştirildi. Her farklı cihaz ile ölçüm arasında 5 dakika ara bırakıldı. Bulgular: Göz içi basıncı ortalamaları GAT'de 14,36±2,50, DKT'de 13,92±2,08 ve NKT'de 9,70±2,81 mmHg olarak bulunmuştur. DKT ile 4 gözde (%6,45), NKT ile de 7 gözde (%11,29) göz içi basıncı ölçülememiştir. Tonometrelerin GAT ile uyumluluk oranı NKT ile %11,30, DKT ile %51,29 olarak bulundu.Sonuç: Keratokonuslu gözlerde GİB ölçümü için NKT'nin uygun olmadığı, GAT ve DKT'nin birbiri ile uyumlu olarak ölçüm yaptıkları tespit edildi.en_US
dc.description.abstractPurpose: To compare the results of intraocular pressure (IOP) measurement of three different tonometers, Goldmann Applanation Tonometry (GAT), Pascal Dynamic Countour Tonometry (DCT) and Keeler Pulsair EasyEye Non-Contact Tonometry (NCT) in keratoconus.Materials and methods: Sixty-two eyes of 36 keratoconus subjects were included in the study. After resting for 5 minutes, patient's IOP was measured. For all subjects, IOP measurements were taken sequentially on the same day by one experienced operator. Three sets of IOP measurements were obtained from each tonometer. Five minutes interval was given between each device. Results: Mean IOPs were 14.36±2.50, 13.92±2.08 and 9.70±2.81 mmHg for GAT, DCT and NCT, respectively. It was impossible to measure IOP in 4 eyes (6.45%) for DCT and in 7 eyes (11.29%) for NCT. The accordance rate was 11.30% between GAT and NCT, and was 51.29% between GAT and DCT.Conclusions: NCT is not reliable for measuring intraocular pressure in keratoconus. We found the accordance between GAT and DCT.en_US
dc.language.isoturen_US
dc.publisherGazi Göz Vakfıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectKeratokonusen_US
dc.subjectGöz İçi Basıncıen_US
dc.subjectGoldmann Aplanasyon Tonometrien_US
dc.subjectPascal Dinamik Kontur Tonometrien_US
dc.subjectNon-Kontakt Tonometrien_US
dc.subjectKeratoconusen_US
dc.subjectIntraocular Pressureen_US
dc.subjectGoldmann Applanation Tonometryen_US
dc.subjectDigital Countour Tonometryen_US
dc.subjectNon-Contact Tonometryen_US
dc.titleKeratokonuslu gözlerde üç farklı tonometre ile göz içi basıncı ölçüm sonuçlarıen_US
dc.title.alternativeIntraocular pressure measurement results with three different tonometers in keratoconusen_US
dc.typearticleen_US
dc.relation.journalGlokom Katarakten_US
dc.departmentHitit Üniversitesi, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümüen_US
dc.identifier.volume12en_US
dc.identifier.issue2en_US
dc.identifier.startpage104en_US
dc.identifier.endpage108en_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster