dc.description.abstractAmaç: Pıhtı aktivatörlü kan alma tüpleri klinik laboratuvarlarda yaygın olarak kullanılmaktadır. Siemens Advia Centaur XP cihazındaki sıra dışı yüksek folat sonuçları kan alma tüplerinin folat ölçümleri üzerine etkisini araştırmamıza sebep oldu. Gereç ve Yöntem: Pıhtı aktivatörlü üç plastik tüp tipi, potansiyel ölçüm interferansı açısından değerlendirildi; Vacutest REF11030 (Grup A), Becton Dickinson (BD) Vacutainer CAT REF367896 (Grup B) ve BD Vacutainer SSTTM II Advance REF367953 (jelli) (Grup C). Görünürde sağlıklı 13 bireyden farklı tüplere alınan kanlarda serum folat konsantrasyonları Siemens Advia Centaur XP, Roche E-170 ve Beckmann Unicell DXI800 cihazlarında ölçüldü ve sonuçlar karşılaştırıldı. İstatistiksel olarak p<0.05 anlamlı kabul edildi. Bulgular: Folat düzeyleri Siemens AdviaCentaur XP cihazı ile ölçüldüğünde; Grup A'da (13.44 ± 4.56 ng/mL) Grup B (8.82 ± 3.33 ng/mL) ve Grup C (8.58 ± 2.83 ng/mL) tüplerinden anlamlı derecede daha yüksekti (p<0.001). Roche E-170 ve Beckmann UniCel DXI800 cihaz sonuçlarında tüp grupları arasında bir farklılık görülmedi (p>0.05). Grup A tüplerine alınmış, farklı cihazlardan elde edilen serum folat düzeyleri karşılaştırıldığında, Siemens Advia Centaur XP cihaz sonuçları (13.44 ± 4.56 ng/mL), Roche E170 (9.15 ± 2.25 ng/mL) ve Beckmann Unicell DXI800 (8.70 ± 3.00 ng/mL) cihaz sonuçlarına göre anlamlı derecede daha yüksekti (p<0.001). Sonuç: Siemens folat kiti, pıhtı aktivatörlü Vacutest tüpünün içeriğinden anlamlı derece etkilenmekte ve sonuçların yüksek çıkmasına sebep olmaktadır. Klasik kalite kontrol programları ile kan alma tüplerinden kaynaklanan ölçüm interferansları tespit edilememektedir. Bundan dolayı her bir yeni kullanıma başlanan kan alma tüpü, ölçüm interferansı açısından özellikle de immünoassay sistemlerinde test edilmelidiren_US
dc.description.abstractObjective: Blood collection tubes with a clot activator are widely used in clinical laboratories. Unexpectedly high results in folate assays obtained on Siemens Advia Centaur XP analyzer triggered a need for an investigation of the effects of blood collection tubes on folate measurements. Material and Methods: We investigated the potential assay interference of three types of plastic tubes with a clot activator; Vacutest REF11030 (Group A); Becton Dickinson (BD) Vacutainer CAT REF367896 (Group B);and BD Vacutainer SSTTM II Advance REF367953 (with gel) (Group C) tubes. Blood samples from apparently healthy 13 volunteers were drawn into these different tube types and serum folate concentrations were analyzed by Siemens Advia Centaur XP, Roche E-170, and Beckmann UniCel DxI800 immunoassay analyzers. The results were compared and p values <0.05 were considered as statistically significant. Results: Folate values measured one Siemens Advia Centaur XP, were significantly higher in Group A (13.44 ± 4.56 ng/mL) than Group B (8.82 ± 3.33 ng/mL) or Group C (8.58 ± 2.83 ng/mL) tubes (p<0.001). No difference was found between the tube groups regarding Roche E-170 and Beckmann UniCel DXI800 analyzers (p>0.05). When serum folate levels drawn into Grup A tubes were compared with different analyzers, results obtained from Siemens Advia Centaur XP (13.44 ± 4.56 ng/mL) were higher than results obtained from Roche E-170 (9.15 ± 2.25 ng/mL) or Beckmann Unicell DXI800 (8.70 ± 3.00 ng/mL) analyzers (p<0.001). Conclusion: Siemens folate reagent was significantly influenced by the contents of the Vacutest tube with a clot activator resulting in high results. Assay interferences born from blood collection tubes represent challenges to clinical laboratories because they are not detected by the usual quality control programs. Therefore, when starting to use a new blood collection tube, assay interferences should be tested, especially for immunoassay systemsen_US
dc.language.isoturen_US
dc.publisherTürk Klinik Biyokimya Derneğien_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectFolaten_US
dc.subjectKan Örneği Alımıen_US
dc.subjectImmunassayen_US
dc.subjectFolateen_US
dc.subjectBlood Specimen Collectionen_US
dc.titleKan alma tüplerinin folat ölçümlerine etkisien_US
dc.title.alternativeThe effects of blood collection tubes on folate measurementsen_US
dc.typearticleen_US
dc.relation.journalTürk Klinik Biyokimya Dergisien_US
dc.departmentHitit Üniversitesi, Tıp Fakültesi, Temel Tıp Bilimleri Bölümüen_US
dc.identifier.volume14en_US
dc.identifier.issue1en_US
dc.identifier.startpage1en_US
dc.identifier.endpage5en_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

Thumbnail

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster