Electrochemical corrosion properties of metal alloys used in orthopaedic implants
Access
info:eu-repo/semantics/closedAccessDate
2009Author
Songür, MuratÇelikkan, Hüseyin
Gökmeşe, Faruk
Şimşek, S. Aykin
Altun, N. Şükrü
Aksu, Mehmet Levent
Metadata
Show full item recordCitation
Songür, M., Çelikkan, H., Gökmeşe, F., Şimşek, S. A., Altun, N. Ş., Aksu, M. L. (2009). Electrochemical corrosion properties of metal alloys used in orthopaedic implants. Journal of Applied Electrochemistry, 39(8), 1259-1265.Abstract
This study concerns an investigation of the corrosion behavior of 316 stainless steel, CoCrMo and Ti6Al4V alloys in simulated body conditions (ringer lactate) at 37 °C by the use of Tafel plots, mixed potential and electrochemical impedance spectroscopy (EIS). Ti6Al4V alloy has the highest corrosion resistance followed by CoCr alloy. Ti6Al4V-CoCrMO was the best couple for galvanic corrosion with the minimum galvanic potential and current values according to mixed potential theory and Tafel method. It was concluded that Ti6Al4V was the most suitable material for implant applications in the human body. © 2009 Springer Science+Business Media B.V.
Source
Journal of Applied ElectrochemistryVolume
39Issue
8Collections
- Makale Koleksiyonu [194]
- Scopus İndeksli Yayınlar Koleksiyonu [2695]
- WoS İndeksli Yayınlar Koleksiyonu [2986]